Ambulatory blood pressure monitoring and Holter ECG

24-hour blood pressure and long-term ECG recording in primary care — indications, set-up, quality requirements and thresholds for interpretation.

Contents (13)

Two recordings that are performed in primary care but interpreted as specialist investigations. Both fail more often for practical than for medical reasons: the cuff is in the wrong place, the electrodes come off, the patient has not kept a diary. The set-up and the instructions decide whether the investigation will be usable.

Ambulatory blood pressure monitoring (ABPM)

Indications

  • Confirmation of newly detected hypertension — recommended before treatment is started.
  • Suspected white-coat hypertension: raised office blood pressure without organ damage.
  • Suspected masked hypertension: normal office blood pressure but organ damage or high risk.
  • Treatment-resistant hypertension.
  • Suspected nocturnal hypertension or absent nocturnal dipping (diabetes, renal failure, sleep apnoea).
  • Suspected hypotension: dizziness, syncope, autonomic dysfunction, adverse drug effects.
  • Wide variability between measurements.
  • Hypertension in pregnancy.

Contraindications

There are no absolute contraindications. Refrain or adapt in atrial fibrillation with a rapid and irregular ventricular rate (oscillometric measurement becomes unreliable), lymphoedema or a dialysis fistula in both arms, an arm circumference too large for the available cuffs, and in a patient who does not tolerate repeated compression (for example marked thrombocytopenia).

Set-up and instructions

  • Choose the cuff according to arm circumference; a cuff that is too narrow overestimates the pressure. Measure the circumference, do not guess.
  • Measure the blood pressure in both arms when fitting the device. Use the arm with the higher pressure, or the non-dominant arm if the values are equal. Check against a manual measurement and document the difference.
  • Programme measurements every 20 minutes during the day and every 30 minutes at night, or alternatively every 15 and every 30 minutes respectively.
  • Instruct the patient to stop, keep the arm still and lowered at each measurement, and not to talk.
  • Otherwise ordinary activities, including work. No showering or bathing.
  • A diary of going to bed, getting up, medication intake, symptoms and any unusual activities.
Upper arm seen from the front with a blood pressure cuff placed in the middle of the upper arm, the tubing led up to a recorder on the belt, and a detail of how far the cuff bladder extends around the arm
Figure 3. The cuff sits in the middle of the upper arm with its lower edge 2–3 cm above the antecubital fossa and the artery marker over the brachial artery on the inner side of the arm. The tubing is led upwards to the recorder, which is attached to the belt. The detail shows the extent of the bladder: it should encircle about 80 percent of the circumference of the arm — a cuff that is too narrow overestimates the pressure.

Interpretation

A recording is acceptable if at least 70 % of the measurements are valid, with at least 20 daytime and 7 night-time values, and with a measurement interval of no more than 30 minutes throughout the 24 hours.

24-hour profile of systolic blood pressure for one patient with normal nocturnal dipping and one without dipping
Figure 1. The 24-hour profile. The green curve shows normal nocturnal dipping: the daytime mean lies around the 135 mmHg threshold and the night-time mean falls by 15 percent to about 112 mmHg, below the night-time threshold of 120 mmHg. The pink curve shows absent dipping (non-dipper) — the night-time values remain at daytime level despite sleep. The sleep period, shaded in the figure, is defined by the patient's diary and never by the clock.
Thresholds for hypertension by method of measurement Office, systolic ≥ 140 mmHg ABPM 24-hour mean, systolic ≥ 130 mmHg ABPM daytime, systolic ≥ 135 mmHg ABPM night-time, systolic ≥ 120 mmHg Home blood pressure, systolic ≥ 135 mmHg The corresponding diastolic thresholds are 90, 80, 85, 70 and 85 mmHg respectively.
Phenotype Office blood pressure ABPM
Normotension Normal Normal
White-coat hypertension Raised Normal
Masked hypertension Normal Raised
Established hypertension Raised Raised

Nocturnal dipping: the pressure normally falls by 10–20 % during the night.

Pattern Nocturnal fall Comment
Dipper 10–20 % Normal
Non-dipper < 10 % Independent risk marker
Reverse dipper ≤ 0 %, night-time pressure higher than daytime The highest risk in the group
Extreme dipper > 20 % Also unfavourable, particularly in the elderly

An abnormal 24-hour rhythm should prompt investigation for sleep apnoea, renal disease and secondary hypertension, and lead to one dose being moved to the evening. The fall can only be calculated if the diary states when the patient fell asleep and got up.

Holter ECG and longer rhythm monitoring

Indications

  • Palpitations, presyncope or syncope of suspected arrhythmic origin.
  • Assessment of rate control in atrial fibrillation.
  • Suspected paroxysmal atrial fibrillation, including after cryptogenic stroke.
  • Evaluation of antiarrhythmic treatment or pacemaker function.
  • Suspected bradyarrhythmia or pauses, particularly at night.
  • Symptom–rhythm correlation in non-specific complaints.

The choice of recording is governed by the frequency of symptoms:

Symptom interval Method
Daily 24-hour Holter
A few times per week 48–72-hour Holter or 7-day recording
About once a month Thumb ECG or event recorder, 2–4 weeks
Less often, with syncope Implantable loop recorder

Contraindications

None. Practical obstacles are skin disease or allergy to the electrode gel, and marked body hair that cannot be shaved.

Set-up

Schematic front view of the torso with electrodes over the clavicles, the sternum and the lower costal margin, and with the muscle bellies to be avoided marked
Figure 2. The principle of electrode placement for Holter monitoring. The electrodes are placed over bony structures — below the clavicles, over the sternum and on the lower costal margin — where the skin moves little relative to the underlying tissue. The hatched areas mark the pectoral muscle and the rectus abdominis, which produce myogenic artefacts as soon as the patient exerts themselves. The number of electrodes and the colour coding differ between manufacturers; the principle does not.
  1. Shave, wash with soap and water, dry and lightly abrade the skin at each electrode site. Grease and skin scales are the commonest cause of unusable recordings.
  2. Place the electrodes over bony structures — sternum, ribs, clavicle — not over muscle bellies, in order to reduce myogenic artefacts.
  3. Relieve strain on the cables with a loop and tape, so that traction does not reach the electrode.
  4. Check the signal quality in all channels before the patient goes home, preferably with the patient sitting, standing and taking deep breaths.
  5. Give a symptom diary and explain the event button: press it at the time of symptoms, note the time and what was being done. Without a diary, symptoms and rhythm cannot be linked.
  6. No showering. Otherwise normal activity.

Interpretation

flowchart TD
  A[The recording read together with the symptom diary] --> B{Did the patient have their typical symptoms during the recording?}
  B -- No --> C[The question is unanswered - extend the recording or change method, do not dismiss the suspicion of arrhythmia]
  B -- Yes --> D{Was there a concurrent arrhythmia at the time of symptoms?}
  D -- Yes --> E[Symptom-rhythm correlation demonstrated - treat the arrhythmia]
  D -- No --> F[A valuable negative finding - argues against an arrhythmic cause, look for another cause]
  A --> G{Findings without concurrent symptoms?}
  G -- Pauses over 3 seconds during the day, Mobitz II or third-degree AV block --> H[Bradyarrhythmia workup, assess for pacemaker indication]
  G -- Episodes of atrial fibrillation --> I[State the AF burden and the ventricular rate, assess the stroke risk]
  G -- PVC burden 10 percent or more, or non-sustained VT --> J[Echocardiography and further cardiological workup]
Finding Comment
Sinus bradycardia down to 30–40/min at night Normal in well-trained and young people
Pauses < 3 seconds at night Often normal; > 3 s during the day requires investigation
Occasional PVCs and PACs Common; a PVC burden ≥ 10 % raises suspicion of PVC-induced cardiomyopathy and warrants echocardiography. The risk rises markedly above 20 %; there is no sharp threshold
Non-sustained VT Investigate for underlying structural heart disease
Episodes of atrial fibrillation State the total AF burden and the ventricular rate
Mobitz II or third-degree AV block Assess for pacemaker indication
Symptoms without concurrent arrhythmia A valuable negative finding — argues against an arrhythmic cause

A Holter ECG without symptoms during the recording does not answer the question. Always state whether the patient had their typical complaints during the recording period.

Complications

Contact dermatitis and skin irritation from the electrodes or the cuff; petechiae and tenderness in the arm after repeated compressions; disturbed sleep from night-time measurements. Serious complications do not occur.

Aftercare and follow-up

Remove the equipment, download the recording, and go through the diary together with the recording — that is where the interpretation is made. Inform the patient of the time to a result.

For ABPM: start or adjust treatment according to the 24-hour values, not according to the office blood pressure. White-coat hypertension is as a rule not treated pharmacologically but must be followed up — a substantial proportion go on to develop established hypertension. Repeat ABPM after 1–2 years.

For Holter: a negative recording with persisting symptoms warrants a longer recording, not dismissal of the question.

Common pitfalls

  • The wrong cuff size — the commonest cause of erroneous ABPM results.
  • No diary. Without sleep times the 24-hour rhythm cannot be calculated, and without notes on symptoms the Holter cannot be correlated.
  • Inadequate skin preparation produces artefacts that make the recording uninterpretable.
  • Too few valid measurements being reported anyway — check the proportion of valid values before the report is written.
  • ABPM in atrial fibrillation gives unreliable oscillometric values.
  • A short Holter recording for infrequent symptoms. Choose the recording time according to the frequency of symptoms.
  • Interpreting night-time values as abnormal without knowing when the patient actually slept.

Authors

EBM AI
Evidensbaserad AI-agent

Updated August 22, 2026